• Cardiac PET
  • HeartSee™
  • Patient Resources
  • Reimbursement
  • Contact
  • Cardiac PET
  • HeartSee™
  • Patient Resources
  • Reimbursement
  • Contact
  • Cardiac PET
  • HeartSee™
  • Patient Resources
  • Reimbursement
  • Contact
  • Cardiac PET
  • HeartSee™
  • Patient Resources
  • Reimbursement
  • Contact
  • Cardiac PET
  • HeartSee™
  • Patient Resources
  • Reimbursement
  • Contact
  • Cardiac PET
  • HeartSee™
  • Patient Resources
  • Reimbursement
  • Contact

Why Bracco?

40 years of excellence in cardiac imaging

Individuals shown are models, not actual patients or healthcare professionals.

Find out why Bracco is the trusted partner in cardiac PET

Bracco is the industry leader in cardiac PET imaging, with more than 3 million diagnostic studies to date. We have helped hundreds of world-class cardiac care facilities, hospitals, and esteemed teaching institutions achieve enhanced diagnostic certainty, operational efficiency, and safety in the diagnosis of coronary artery disease (CAD).

The Bracco difference:

  • An extensive team of accessible, locally based experts stand ready to support you and your staff
  • Exclusive, customized business case tools help ensure you meet your financial goals
  • Drive clinical and operational excellence in your organization with Bracco’s peer-to-peer support program
  • Ongoing, complimentary clinical training at your facility keeps your staff at its best
  • Bracco supports the business side of your practice with personalized reimbursement training and updates
  • Flexible generator delivery schedules make cardiac PET imaging available to all—regardless of patient volume, regardless of location
Explore Our Mobile Isotope Service

Bracco was the first company to drive the adoption of cardiac PET through innovation, research, and development. Today, we are committed to growth and preservation of the modality for you and your patients.

Resources

For more information about nuclear medicine, radiology, healthcare administration and policy, cardiology, accreditation and certification, and more, see the links below.

This Internet site may provide links or references to other sites but the owners of this site have no responsibility for the content of such other sites and shall not be liable for any damages or injury arising from that content. Any links to other sites are provided as merely a convenience to the users of this Internet site.

  • American Society of Nuclear Cardiology (ASNC)
  • Society of Nuclear Medicine and Molecular Imaging (SNMMI)
  • Medical Imaging and Technology Alliance (MITA)
  • MedAxiom
  • American College of Cardiology (ACC)

About Bracco

Bracco group, founded in 1927, is a global leader in diagnostic imaging, committed to advancing healthcare and improving people’s lives by shaping the future of prevention and precision medicine.

The company operates in the healthcare sector across more than 100 countries with a workforce of over 4,000 employees and consolidated annual revenues of approximately €2 billion, 88% generated by international markets.

With a strong commitment to innovation—investing around 9% of its reference turnover in Research & Development—Bracco develops and provides a broad portfolio of pharmaceutical products for diagnostic imaging, including contrast agents for X-ray, Computed Tomography (CT), and Magnetic Resonance Imaging (MRI), as well as microbubbles for Contrast Enhanced Ultrasound (CEUS), and Molecular Imaging through radioactive tracers and novel PET imaging agents, alongside AI-based solutions. It is also a global market leader in advanced contrast management technologies for cardiovascular angiography and radiology imaging.

Discover More

CARDIOGEN-82®
(rubidium Rb 82 generator)

INDICATION

CARDIOGEN-82 is a closed system used to produce rubidium Rb 82 chloride injection for intravenous administration and is indicated for Positron Emission Tomography (PET) imaging of the myocardium under rest or pharmacologic stress conditions to evaluate regional myocardial perfusion in adult patients with suspected or existing coronary artery disease.

IMPORTANT SAFETY INFORMATION

WARNING: HIGH LEVEL RADIATION EXPOSURE WITH USE OF INCORRECT ELUENT AND FAILURE TO FOLLOW THE ELUATE TESTING PROTOCOL

High Level Radiation Exposure with Use of Incorrect Eluent

Patients are exposed to high radiation levels when the CardioGen-82 generator is eluted with the incorrect eluent due to high Sr 82 and Sr 85 breakthrough levels.

  • Use only additive-free 0.9% Sodium Chloride Injection USP to elute the generator
  • If an incorrect solution is used to elute CardioGen-82:
    • Immediately stop the patient infusion;
    • Evaluate the patient’s radiation absorbed dose and monitor for the effects of radiation to critical organs such as bone marrow;
    • Permanently discontinue use of the affected generator.

Excess Radiation Exposure with Failure to Follow the Eluate Testing Protocol

Excess radiation exposure occurs when the levels of Sr 82 or Sr 85 in the rubidium Rb 82 chloride injection exceed specified limits.

  • Record each generator eluate volume, including waste and test volumes, and keep a record of the cumulative eluate volume
  • Strictly adhere to the generator eluate testing protocol, to minimize the risk of excess radiation exposure, including daily testing and additional testing at Alert Limits
  • Stop using the generator if it reaches any of its Expiration Limits: 17 L for the generator’s cumulative eluate volume, or 42 days post generator calibration date, or an eluate Sr 82 level of 0.01 μCi /mCi Rb 82, or an eluate Sr 85 level of 0.1 μCi /mCi Rb 82

WARNINGS AND PRECAUTIONS

High Level Radiation Exposure with Use of Incorrect Eluent

Use only additive free 0.9% Sodium Chloride Injection USP to elute the generator. Apply the provided saline tag to the additive free 0.9% Sodium Chloride Injection USP container before use. Additives present in other solutions (particularly calcium ions) expose patients to high levels of radiation by causing the release of large amounts of Sr 82 and Sr 85 into the eluate regardless of the generator’s age or prior use. Immediately stop the patient infusion and discontinue use of the affected CardioGen-82 generator if the incorrect eluent is used and evaluate the patient’s radiation absorbed dose and monitor for the effects of radiation to critical organs such as bone marrow.

Excess Radiation Exposure with Failure to Follow Testing Protocol

Excess radiation exposure occurs when the Sr 82 and Sr 85 levels in rubidium Rb 82 chloride injections exceed the specified generator eluate limits. Strictly adhere to the eluate testing protocol to minimize radiation exposure to the patient. Stop using the rubidium generator when the expiration limits are reached.

Risk Associated with Pharmacologic Stress

Pharmacologic induction of cardiovascular stress may be associated with serious adverse reactions such as myocardial infarction, arrhythmia, hypotension, bronchoconstriction, and cerebrovascular events. Perform pharmacologic stress testing in accordance with the pharmacologic stress agent’s prescribing information and only in the setting where cardiac resuscitation equipment and trained staff are readily available.

Cumulative Radiation Exposure: Long-Term Risk of Cancer

Rubidium Rb 82 chloride injection, similar to other radiopharmaceuticals, contributes to a patient’s overall long-term cumulative radiation exposure which is associated with an increased risk of cancer. Use the lowest dose of rubidium Rb 82 chloride injection necessary for imaging and ensure safe handling to protect the patient and health care worker. Encourage patients to void as soon as a study is completed and as often as possible thereafter for at least one hour.

ADVERSE REACTIONS

Radiation Exposure

High level radiation exposure to the bone marrow has occurred in some patients due to Sr 82 and Sr 85 breakthrough in the eluate when an incorrect solution was used to elute the rubidium Rb 82 generator. Excess radiation exposure has occurred in some patients who received rubidium Rb 82 chloride injections at clinical sites where generator eluate testing appeared insufficient.

You are encouraged to report negative side effects of prescription drugs to the FDA. Visit www.fda.gov/medwatch or call 1-800-FDA-1088.

Please click here for full Prescribing Information for CardioGen-82® (rubidium Rb 82 generator) including BOXED WARNING.

CARDIOGEN-82 is manufactured for Bracco Diagnostics Inc., Princeton, NJ 08540, by GE Healthcare, Medi-Physics, Inc., Arlington Heights, IL 60004.

CARDIOGEN-82 is a registered trademark of Bracco Diagnostics Inc.

WE ARE CARDIAC PET is a registered trademark of Bracco Diagnostics Inc.

Bracco Diagnostics Inc.
510 Carnegie Center
Suite 300
Princeton, NJ 08540 USA

Phone: 609-514-2200
Toll-Free: 1-800-631-5245 (U.S. only)
Fax: 609-514-2424

WE ARE CARDIAC PET is a registered trademark of Bracco Diagnostics Inc.
©2026 Bracco Diagnostics Inc.
All Rights Reserved. US-CG-2000009 09/26

  • Privacy Policy
  • Terms of Use
  • Privacy Policy
  • Terms of Use

Click HERE to subscribe to the MedAxiom
HeartTalk Podcast Series.

We look forward to sharing valuable insights and information with you. 

Please provide your information below and check your email for a confirmation.

"*" indicates required fields

This field is for validation purposes and should be left unchanged.
Name*
*